Commands Listed in Alphabetical Order
Syntax
CALibrate:UPD
ATEinfo:MAInframe
Examples
CALIBRATE:UPDATEINFO:MAINF RAME u
pdates the nonvolatile memory of the
mainframe with the latest calibration information.
CH<x>? (Query Only)
This is a query only command that returns the vertical parameters for the specified
channel. The channel is specified by x, which can be 1 through 8.
Group
Vertical
Syntax
CH<x>?
Examples
CH3? might r eturn CH3:DESKEW 0.000000000000; OFFSET
0.000000000000; UNITS AUTO; FILTER:VALUE NONE;
CH3:EXTATTEN:MODE LINEAR; VALUE 1.000000000000; CH3:POSITION
0.000000000000; PTPEAK 0.700000000000; SCALE 0.700000000000
.
CH<x>:BANdwidth:LIST? (Query Only)
This query-only command returns a list of the available bandwidths for the
specified channel. The channel is specified by x, which can be 1 through 8.
The returned list contains the actual bandwidths available for the channel. The
available bandwidths are dependent on the module.
The optical modules have differing valid settings. Use the <>:LIST? query to
determine the valid {Clock Recovery | Filter | Bandwidth | Wavelength} settings
for your optical module. Features of currently installed optical modules can be
viewed using the System Properties dialog.
To set the bandwidth for a channel, use the CH<x>:BANdwidth:VALue command.
Group
Vertical
Syntax
CH<x>:BANdwidth:LIST?
Related Commands
CH<x>:BANdwidth:VALue
DSA/CSA/TDS8X00/B Series Programmer Manual 2-77