Commands Listed in Alphabetical Order
Arguments
Floating point
number that represents the AFG phase, in degrees.
Examples
AFG:PHASE -145.0 sets the phase of the AFG to -145.0 degrees.
AFG:PULse:WIDth
Sets (or queries) the AFG pulse width, in seconds.
Pulse width has an absolute minimum of 10ns and has a relative range of 10% -
90% of the current period setting. Resolution is 0.1ns.
Conditions
MDO3000 series with option MDO3AFG installed and MDO4000C series with
option MDO4AFG installed only.
Group
AFG
Syntax
AFG:PULse:WIDth <NR3>
AFG:PULse:WIDth?
Arguments
Floating point number that represents the pulse width, in seconds.
Examples
AFG:PULSE:WIDTH 100.0E-6 sets the AFG pulse width to 100 microseconds.
AFG:RAMP:SYMmetry
Sets (or queries) the AFG ramp symmetry as a percentage. Minimum is 0.0%,
maximum is 100.0% and increment is 0.10%.
Conditions
MDO3000 series with option MDO3AFG installed and MDO4000C series with
option MDO4AFG installed only.
Group
AFG
Syntax
AFG:RAMP:SYMmetry <NR3>
AFG:RAMP:SYMmetry?
Arguments
Floating point number that represents the AFG ramp symmetry, as a percentage.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-141