Commands Listed in Alphabetical Order
AUXin:PRObe:
AUTOZero (No Query Form)
This command executes the attached probe’s Auto Zero function, for probes that
support this feature. See your probe documentation for more details.
Conditions
Available for MSO/DPO4000B series models, 2-channel MDO3000 models, and
MDO4000C models without option SA3 or SA6 only.
Group
Vertical
Syntax
AUXin:PRObe:AUTOZero EXECute
Arguments
EXECute sets the probe to autozero.
Examples
AUXin:PRObe:AUTOZero EXECute performs the AutoZero function in the
probe.
AUXin:PRObe:CALibrate:CALIBRATABLe? (Query Only)
This query indicates whethe
r the attached probe is calibratable. It returns a
boolean value.
Conditions
Available for MSO/DPO4000B series models, 2-channel MDO3000 models, and
MDO4000C models without option SA3 or SA6 only.
Group
Vertical
Syntax
AUXin:PRObe:CALibrate:CALIBRATABLe?
Examples
AUXin:PRObe:CALibrate:CALIBRATABLe? might return 0, indicating that the
probe attached to the Aux Input connector is not calibratable.
AUXin:PRObe:COMMAND (No Query Form)
Sets the state of the probe control specified with the first argument to the state
specified with the second argument. The commands and states are unique to the
attached probe type. Only certain VPI probes support this command. See the
probe documentation for how to set these string arguments.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-151