Commands Listed in Alphabetical Order
RECOMMENDED in
dicates the measurement accuracy might be improved by
degaussing the probe.
PASSED indica
tes the probe is degaussed.
FAILED indicates the degauss operation failed.
RUNNING indicates the probe degauss operation is currently in progress.
CH<x>:PRObe:FORCEDRange
This comma
nd specifies the range of a TekVPI probe attached to the channel
specified by <x>, where x is the channel number.
NOTE. This command will return
PASSED
for probes that do not support degauss
operations.
Group
Vertical
Syntax
CH<x>:PRObe:FORCEDRange <NR3>
CH<x>:PRObe:FORCEDRange?
Arguments
<NR3> is a floating point number that specifies the range, which is probe specific.
Returns
The query form of this command returns 0.0 for probes that do not support
forced ranges.
CH<x>:PRObe:GAIN
This command specifies the gain factor for the probe attached to the channel
specified by <x>, where x is the channel number. The "gain" of a probe is the
output divided by the input transfer ratio. For example, a common 10x probe
has a gain of 1V.
Group
Vertical
Syntax
CH<x>:PRObe:GAIN <NR3>
CH<x>:PRObe:GAIN?
Related Commands
CH<x>:SCAle
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-229