Commands Listed in Alphabetical Order
Syntax
DISplay:INTEN
SITy:BACKLight:AUTODim:ENAble {OFF|ON|0|1}
DISplay:INTENSITy:BACKLight:AUTODim:EN Able?
Arguments
OFF or 0 turns off the display auto-dim feature.
ON or 1 turns it on.
DISplay:INTENSITy:BACKLight:AUTODim:TIMe
Sets or returns the amount of time, in minutes, to wait for no UI activity before
automatically dimming the display. The time can range from a minimum of
10 minutes to a maximum of 1440 minutes (24 hours). The default is 60 minutes.
Group
Display
Syntax
DISplay:INTENSITy:BACKLight:AUTODim:TI Me <NR1>
DISplay:INTENSITy:BACKLight:AUTODim:TI Me?
Arguments
An integer that represents the amount of time, in minutes, to wait for no UI
activity before automatically dimming the display.
DISplay:INTENSITy:GRAticule
Sets and ret
urns the display graticule intensity settings.
Group
Display
Syntax
DISplay:INTENSITy:GRAticule <NR1>
DISplay:INTENSITy:GRAticule?
Arguments
<NR1> is the graticule intensity and ranges from 0 t
o100percent.
Examples
DISPLAY:INTENSITY:GRATICULE? might return
DISPLAY:INTEnsITY:GRATICULE 30
DISplay:INTENSITy:WAVEform
Sets and returns the display waveform intensity settings.
2-292 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual