Commands Listed in Alphabetical Order
Arguments
1 or ON turns on t
he auto range state for the Digital Voltmeter.
0 or OFF turns it off.
Examples
DVM:AUTOR ON turns on the auto range state for the Digital Voltmeter.
DVM:DISPLAYSTYle
Sets (or queries) the display style for the Digital V
oltmeter.
Conditions
MDO3000 and MDO4000C only; requires DVM option key to be installed.
Group
DVM
Syntax
DVM:DISPLAYSTYle {FULl|MINimum}
DVM:DISPLAYSTYle?
Arguments
FULl sets the display style for the DVM to Full.
MINimum sets the display style for the DVM to Minimum.
Examples
DVM:DISPLAYSTYLE MIN sets the display style for the DVM to Minimum.
DVM:MEASUrement:FREQuency? (Query Only)
Returns the current frequency value for the Digital Voltmeter.
Any qualifier, such as "Measurement warning, Clipping positive", is automatically
set as an error event.
Conditions
MDO3000 and MDO4000C only; requires DVM option key to be installed.
Group
DVM
Syntax
DVM:MEASUrement:FREQuency?
Examples
DVM:MEASU:FREQ? might return 100.0000 E+3, which represents 100 KHz as
the current frequency value for the DVM.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-297

Table of Contents

Other manuals for Tektronix MDO3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO3000 Series and is the answer not in the manual?

Tektronix MDO3000 Series Specifications

General IconGeneral
BrandTektronix
ModelMDO3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals