Commands Listed in Alphabetical Order
ACDCRMS –displ
ays the RMS value of the acquired data.
DC – displays the DC value of the acquired data.
FREQuency – displays the frequency counter measurement.
OFF
Examples
DVM:MODE DC sets the mode for the DVM to DC, which displays the DC value of
the acquired data.
DVM:MOD? might return ACRMS, which indicates the mode is currently set
to ACRMS.
DVM:SOUrce
Sets (or queries) the source for the Digital Voltmeter: Channel 1 - 4.
Conditions
MDO3000 and MDO4000C only; requires DVM option key to be installed.
Group
DVM
Syntax
DVM:SOUrce {CH1|CH2|CH3|CH4}
DVM:SOUrce?
Arguments
CH1|CH2|CH3|CH4 specify which channel to use as the source for the DVM.
Examples
DVM:SOURCE CH4 sets the DVM source to Channel 4.
D<x>
This command specifies parameters for digital channel <x>, where x is the
channel number.
Conditions
Requires an MSO or MDO4000/B/C model, or MDO3000 with option
MDO3MSO installed.
Group
Vertical
Syntax
D<x>
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-301