Commands Listed in Alphabetical Order
Arguments
NEXT moves to th
e next reference mark on the right.
PREVious moves to the next reference mark on the left.
MARK:CREATE
(No Query Form)
Creates a mark on a specified waveform or all waveforms in a column.
Group
Mark
Syntax
MARK:CREATE {CH1|CH2|CH3|CH4|MATH|REF1
|REF2|REF3|REF4|B1|B2|B3|B4|REF1|REF2| REF3|REF4|DIGital|
COLUMN|RF_AMPlitude|RF_FREQuency|RF_PH ASe}
Arguments
CH<x> creates the mark on a channel waveform, where <x> is the channel number.
MATH creates the mark
on the math waveform.
B<x> creates the mark on a bus waveform, where <x>
REF<x> creates the m
ark on a reference waveform, where <x> is the reference
waveform number.
DIGital creates th
e mark on a digital waveform. (An error will result if no digital
channel is turned on.) (MSO/MDO4000/B models only as well as MDO3000 and
MDO400C models with option MDO3MSO or MDO4MSO installed.)
COLUMN creates marks on all waveforms in the current zoom pixel column.
RF_AMPlitude|R
F_FREQuency|RF_PHASe
creates the mark on an RF time
domain trace. (MDO4000/B/C models only.)
MARK:DELEte (No Query Form)
This command deletes a mark on a particular waveform, all waveforms in a
column, the selected mark, or all marks.
Group
Mark
Syntax
MARK:DELEte
{CH1|CH2|CH3|CH4|MATH|REF1|REF2|REF3|R EF4|B1|B2|B3|B4|REF1|REF2|
REF3|REF4|DIGital|COLUMN|SELE
CTED|ALL||RF_AMPlitude|RF_FREQuency|
RF_PHASe}
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-353