Commands Listed in Alphabetical Order
Syntax
MEASUrement:M
ETHod {Aut o|HIStogram|MINMax}
MEASUrement:METHod?
Related Commands
MEASUrement:REFLevel:PERCent:HIGH,
MEASUrement:REFLevel:PERCent:LOW,
MEASUrement:REFLevel:PERCent:MID<x>,
Arguments
Auto selects the best method for each data set.
HIStogram sets the high and low waveform levels statistically using a histogram
algorithm.
MINMax uses the highest and lowest values of the waveform record. This selection
is best for examining waveforms with no large, flat portions of a common value
,
such as sine waves and triangle waves.
Examples
MEASUREMENT:METHOD? might return :MEASUREMENT:METHOD MINMAX
indicating that the reference levels are set to MIN and MAX.
MEASUrement:REFLevel? (Query Only)
Returns the current reference level parameters.
Group
Measurement
Syntax
MEASUrement:REFLevel?
Examples
MEASUREMENT:REFLEVEL? might return these reference
level settings
:MEASUREMENT:REFLEVEL:METHOD
PERCENT;ABSOLUTE:HIGH 0.0000;LOW 0.0000;MID1 0.0000 ;MID2
0.0000;:MEASUREMENT:REFLEVEL:PERCENT:H IGH 90.0000;LOW
10.0000;MID1 50.0000 ;MID2 50.0 000
MEASUrement:REFLevel:ABSolute:HIGH
This command specifies the high reference level, and is the upper reference level
when MEASUrement:REFLevel:METHod is set to Absolute. This command
affects the results of rise and fall measurements.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-445