Commands Listed in Alphabetical Order
Group
Power
Syntax
POWer:HARMonics:RESults:IEC:POHL?
Examples
POWER:HARMONICS:RESULTS:IEC:POHL? might return
:POWER:HARMONICS:RESULTS:IEC:POHL 0.5 indicating that the
IEC POHL is set to 0.5 A.
POWer:HARMonics:RESults:IEC:POWer? (Query Only)
Returns the IEC input power that is used to calculate limits.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:HARMonics:RESults:IEC:POWer?
Examples
POWER:HARMONICS:RESULTS:IEC:POWER? might return
:POWER:HARMONICS:RESULTS:IEC:POWER 5 indicating that the
IEC power is set to 5 W.
POWer:HARMonics:RESults:IEC:POWERFactor? (Query Only)
Returns the IEC power factor measurement.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:HARMonics:RESults:IEC:POWERFacto r?
Examples
POWER:HARMONICS:RESULTS:IEC:POWERFACTO R? might return
:POWER:HARMONICS:RESULTS:IEC:POWERFACT OR 9.1900E+37 indicating
that the IEC power factor is set to 9.1900E+37.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-483

Table of Contents