Commands Listed in Alphabetical Order
POWer:RIPPle
:SOUrce
This command specifies the source waveform for ripple tests. The voltage source
waveform is specified using the POWer:VOLTAGESOurce command and the
current wave
form is specified using the POWer:CURRENTSOurce command.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:RIPPle:SOUrce {VOLTage|CURRent}
Arguments
VOLTage specifies voltage source waveform for ripple tests.
CURRent specifies current source waveform for ripple tests.
Examples
POWER:RIPPLE:SOURCE CURRent sets the ripple source to Current.
POWER:RIPPLE:SOURCE? might return :POWER:RIPPLE:SOURCE VOLT
indicating that the ripple source is set to Voltage.
POWer:SOA:LINear:XMAX
This command specifies the user XMAX value for use in linear SOA calculations.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SOA:LINear:XMAX <NR3>
POWer:SOA:LINear:XMAX?
Arguments
<NR3> is a floating point number that specifies the XMAX value used for linear
SOA calculations.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-507