Commands Listed in Alphabetical Order
Examples
POWER:SOA:LOG
:XMAX 1
sets the XMAX value to 1.0000.
POWER:SOA:LOG:XMAX ? might return :POWER:SOA:LOG:XMAX 1.0000E+3
indicating th
at the XMAX value for log plot type is set to 1.0000E+3.
POWer:SOA:LOG:XMIN
This command specifies the user XMIN value for use in Log SOA calculations.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SOA:LOG:XMIN <NR3>
POWer:SOA:LOG:XMIN?
Arguments
<NR3> is a floating point number that specifies the XMIN value used for log
SOA calculations.
Examples
POWER:SOA:LOG:XMIN 2 sets the XMIN value to 2.0000.
POWER:SOA:LOG:XMIN ? might return :POWER:SOA:LOG:XMIN 100.0000E-3
indicating that the XMIN value for log plot type is set to 100.0000E-3.
POWer:SOA:LOG:YMAX
This command specifies the user YMAX value for use in Log SOA calculations.
Conditions
This command requires a DPO4PWR application module for 4000 series models,
or an MDO3PWR application module for MDO3000 series models.
Group
Power
Syntax
POWer:SOA:LOG:YMAX <NR3>
POWer:SOA:LOG:YMAX?
Arguments
<NR3> is a floating point number that specifies the YMAX value used for log
SOA calculations.
2-510 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual