Commands Listed in Alphabetical Order
Arguments
LANdscape spec
ifies that screen images are saved in landscape format.
PORTRait specifies that screen images are saved in portrait format.
SAVe:MASK (No Query Form)
This command saves the current mask definition to the file specified with a quoted
string, into the current working directory. You can recall the mask from thefileby
using the command RECAll:MASK
NOTE. If you would like to save the mask denition plus the rest of the scope
settings to the current working directory, use the command
SAVE:SETUP .
A series of examples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Example
s.)
Conditions
This command requires a DPO4LMT application module.
Group
Mask
Syntax
SAVe:MASK <QString>
Related Commands
RECAll:MASK
Arguments
QString is a quoted string that is the name of the mask definition being saved to
the current working directory.
Examples
:SAVE:MASK ”FOO.MSK” will save a copy of the mask “FOO” to the current
working directory.
SAVe:SETUp (No Query Form)
Stores the state of the oscilloscope to a specified memory location. You can later
use the *RCL command to restore the oscilloscope to this saved state.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-597