Commands Listed in Alphabetical Order
TRIGger:A:BU
S:B<x>:CAN{:IDentifier|:ADDRess}:VALue
This command sets the binary address value to be used when triggering on a CAN
bus signal. The trigger condition must be set to IDANDDATA OR DATA (using
TRIGger:A:B
US:B<x>:CAN:CONDition).
B<x> is the bus number (1-4). 1-3 for
the MDO4000C and 1-2 for the MDO3000.
Conditions
This comm
and requires a DPO4AUTO or DPO4AUTOMAX application module
on DPO/MSO4000B and MDO4000/B/C models and a MDO3AUTO application
module on MDO3000 models.
Group
Trigger
Synt
ax
TRIGger:A:BUS:B<x>:CAN{:IDentifier|:AD DRess}:VALue <QString>
TRIGger:A:BUS:B<x>:CAN{:IDentifier|:AD DRess}:VALue?
Arguments
<QString> is up to 29 bits specifying the binary identifier value. The only
allowedcharactersinthe
QString are 0, 1, and X.
Examples
TRIGGER:A:BUS:B1:CAN:IDENTIFIER:VALUE 1011 sets the identifier value
to 1011.
TRIGger:A:BUS:B<x>:ETHERnet:CONDition
This command specifies a field or condition within an Ethernet frame to trigger
on.
B<x> is the bus number B1–B4.
Conditions
This command requires a DPO4ENET application module for DPO/MSO4000B
and MDO4000/B/C models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:ETHERnet:CONDition
{SFD|MACADDRess|MACLENgth|IPHeader|TCP Header|DATa|EOP|IDLe
|FCSError|QTAG}
TRIGger:A:BUS:B<x>:ETHERnet:CONDition?
Related Commands
Most of the other TRIGger:A:BUS:B <x>:ETHERnet commands are impacted
by the setting of this command.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-793