Commands Listed in Alphabetical Order
Conditions
This command re
quires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:MIL1553B:STATus:PAR ity
{0|1|X|ZERo|ONE|NOCARE|OFF|ON}
TRIGger:A:BUS:B<x>:MIL1553B:STATus:PAR ity?
Related Commands
TRIGger:A:BUS:B<x>:MIL1553B:CONDition
Arguments
0
1
X
sets the value to X (“don’t care”) which is the default.
ZERO sets the value to 0.
ONE sets the value to 1.
NOCARE sets the value to X (“don’t care”) which is the default.
OFF sets the value to 0.
ON sets the value to 1.
Examples
TRIGger:A:BUS:B1:MIL1553B:STAT:PAR ONE sets the parity bit value to be
used in the trigger to 1.
TRIG:A:BUS:B1:MIL1553B:STATUS:PARity? might return X, indicating
that the parity doesn’t matter.
TRIGger:A:BUS:B<x>:MIL1553B:TIMe:LESSLimit
When the MIL-STD-1553 bus trigger condition is set to TIMe, this command
specifies either the minimum remote terminal response time (RT) limit
for the amount of time the terminal has to transmit, or it specifies the
minimum inter-message gap (IMG). (You can specify RT or IMG using the
TRIGger:A:BUS:B<x>:MIL1553B:CONDition TIMe command.)
B<x>
Conditions
This command requires a DPO4AERO application module for DPO/MSO4000B
and MDO4000/B/C models or MDO3AERO application module for
MDO3000 models.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-851