Commands Listed in Alphabetical Order
TRIGger:A:SE
THold:CLOCk:SOUrce
This command specifies the clock source for the setup and hold triggering. You
cannot specify the same source for both clock and data.
Group
Trigger
Syntax
TRIGger:A:SETHold:CLOCk:SOUrce
{CH1|CH2|CH3|CH4|D0|D1|D2|D3|D4|D5|D6| D7|D8|D9|D10|D11|D12
|D13|D14|D15|AUX}
TRIGger:A:SETHold:CLOCk:SOUrce?
Related Commands
TRIGger:A:SETHold:DATa:SOUrce
Arguments
CH1–CH4 specifies the analog channel to use as the clock source waveform.
D0–D15 specifies the digital channel to use as the clock source waveform.
(MSO/MDO4000/B models only as well as MDO3000 and MDO400C models
with option MDO3MSO or MDO4MSO installed.)
AUX specifies an external trigger using the Aux In connector located on the front
panel of the oscilloscope. (For MSO/DPO 4000B series models, MDO4000C
series models without an RF option, and 2-channel MDO3000 models only.)
Examples
TRIGGER:A:SETHOLD:CLOCK:SOURCE CH1 specifies channel 1 as the clock
source for a setup and hold trigger operation.
TRIGGER:A:SETHOLD:CLOCK:SOURCE? might return :TRIGG ER:A:SETHOLD:
CLOCK:SOURCE CH4
indicating that channel 4 is the clock source for a setup and
hold trigger operation.
TRIGger:A:SETHold:CLOCk:THReshold
This command specifies the clock voltage threshold for the setup and hold trigger.
Group
Trigger
Syntax
TRIGger:A:SETHold:CLOCk:THReshold {<NR3>|TTL}
TRIGger:A:SETHold:CLOCk:THReshold?
2-900 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual