Commands Listed in Alphabetical Order
Group
Trigger
Syntax
TRIGger:A:VIDeo:CUSTom:LINEPeriod <NR3>
TRIGger:A:VIDeo:CUSTom:LINEPeriod?
Arguments
<NR3> is the custom video line period.
Examples
TRIGGER:A:VIDEO:CUSTOM:LINEPERIOD 50.5E-6 sets the video line period
to 50.5 μs.
TRIGGER:A:VIDEO:CUSTOM:LINEPERIOD? might return
TRIGGER:A:VIDEO:CUSTOM:LINEPERIOD 63.5600E-6 indicating
the video line period is set to 63.56 μs.
TRIGger:A:VIDeo:CUSTom:SYNCInterval
This command sets the sync interval for the standard under test to use for
triggering on video signals. This is only required for BiLevel Custom. To use
this command, you must also set the video standard to
BILevelcustom (using
TRIGger:A:VIDeo:STANdard).
Conditions
This command requires a DPO4VID application module to use any standard
besides NTSc, PAL, or SECAM, except for MDO3000 models, which do not
require an application module.
Group
Trigger
Syntax
TRIGger:A:VIDeo:CUSTom:SYNCInterval <NR3>
TRIGger:A:VIDeo:CUSTom:SYNCInterval?
Arguments
<NR3> is the sync interval.
Examples
TRIGGER:A:VIDEO:CUSTOM:SYNCINTERVAL 4.0E-6 sets the sync interval is
set to 4.0 μs.
TRIGGER:A:VIDEO:CUSTOM:SYNCINTERVAL? might return
TRIGGER:A:VIDEO:CUSTOM:SYNCINTERVAL 4.7200E-6 indicating the sync
interval is set to 4.72 μs.
2-914 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual