Pinpoint Trigge
rs
1. To more accu rat
ely place the trigger
on the display, check Enhanced
Triggering. To more accurately place
the trigger on n
oisy signals, also check
Optimized for Noisy Signals .
The Optimized for Noisy Signals
selection is o
nly available if Enhanced
Triggering is selected.
Triggering with B-Event Scan
Use A->B Sequence B-Event Scan to create overlapped eye diagrams synchronized or initiated by an A trigger event.
Trig on nth Event captures the nth B-Event that occurs after the A-Event for all acquisitions. B-Event Scan automatically
advances the B -Event value to capture different portions of the signal.
1. Set the A trigger type and source in the
A Event tab.
2. Set the
B trigger type and source in the
B Event tab.
MSO70000/C, DPO/DSA70000B/C, DPO7000, and MSO/DP O5000 Series U ser M anual 77