Performance Tests (MSO/DPO5000 Series)
8. Record the nega
tive-measured and positive-measured mean readings in the
worksheet as follows:
a. Set the DC Volt
ageSourcetoV
negative
.
b. Push the front-panel Wave Inspector Measure button.
c. In the Measurements Setup window, go to the Measurements box. Click
the More button and then select Mean.
d. View the Mean measurement value in the display.
e. In the Measurement Setup window, select Statistics.
f. Click Reset.
g. Enter the mean reading in the worksheet as V
negative-measured
.(SeeTable4-2
on page 4-32.)
h. Set the DC Voltage Source to V
positive
.
i. Click R
eset again.
j. Enter the mean reading in the worksheet as V
positive-measured
.
MSO70000/C, DSA70000B/C, DPO7000B/C, DPO7000, MSO5000, DPO5000 Series 4–31