Performance Tests (MSO/DPO5000 Series)
4. Record the Low a
nd High measurements at 1 MΩ as follows:
a. Push the front-panel Wave Inspector Measure button.
b. Ensure that the Amplitude tab is selected.
c. Select the Low measurement.
d. Enter the Low measurement reading in the test record.
e. Select the High measurement.
f. Enter the H
igh measurement reading in the test record.
5. Record the Low and High measurements at 50 Ω as follows:
a. Push the front-panel oscilloscope Vertical Menu button.
b. Set the Termination to 50 Ω.
c. Push the fr ont-panel Wave Inspector Measure button.
d. Ensure that the Amplitude tab is selected.
e. Select
the Low measurement.
f. Ente r the Low measurement reading in the test record.
g. Select the High measurement.
h. Enter the High measurement reading in the test record.
This completes the performance verification procedure.
4–48 MSO70000/C, DSA70000B/C, DPO7000B/C, DPO7000, MSO5000, DPO5000 Series