EasyManuals Logo

Tektronix TDS 200-Series Service Manual

Tektronix TDS 200-Series
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #45 background imageLoading...
Page #45 background image
Performance Verification
TDS 200 Series Digital Oscilloscope Service Manual
49
TDS 210 and TDS 220 only: this test checks the edge trigger sensitivity for the
external trigger.
1. Set up the instrument using the following steps:
Press menu button Select menu item Select setting
SAVE/
RECALL
Recall
Factory
CH 1 Probe 1X
TRIGGER
Source Ext
Mode Normal
ACQUIRE Sample
MEASURE
Source CH1
Type Pk-Pk
2. Connect the instrument to the leveled sine wave generator as shown below.
Digitizing oscilloscope
Leveled
sine wave
generator
Output
50 W
feedthrough
terminator
BNC
cable
BNC cable
BNC T connector
3. Set the leveled sine wave generator frequency to 60 MHz if you are checking
a TDS 210 or to 100 MHz if you are checking a TDS 220.
4. Set the instrument VOLTS/DIV to 500 mV/div.
5. Set the instrument SEC/DIV to 10 ns/div.
6. Set the leveled sine wave generator output level to approximately 750 mV
p-p
so that the measured amplitude is approximately 750 mV. (The measured
amplitude can fluctuate around 750 mV.)
7. Press SET LEVEL TO 50%. Adjust TRIGGER LEVEL as necessary and
then check that triggering is stable.
Check External Edge
Trigger Sensitivity

Table of Contents

Other manuals for Tektronix TDS 200-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 200-Series and is the answer not in the manual?

Tektronix TDS 200-Series Specifications

General IconGeneral
BrandTektronix
ModelTDS 200-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals