11. Send the command MEASUrement:IMMed:SOURCE CH1 to
select measurements on channel 1.
12. Send the command MEASUrement:IMMed:TYPe PK2 to set up
the voltage measurement.
13. Send the query MEASUrement:IMMed:VALue? to request the
measurement result. The oscilloscope will respond with a result
similar to , which is the voltage measurement of the
PROBE COMP signal using the standard 10x probe.
This completes the RS-232 interface test.