TDS 420A, TDS 430A, TDS 460A & TDS 510A User Manual
3–97
Remote Communication
This section explains how to prepare and setup the oscilloscope for control and
operation over the IEEE Std 488.2-1987 (GPIB) interface. For more information,
see your programmer manual and the TDS Family Option 13 Instruction Manual
(Option 13 equipped instruments only).
To set up remote communications, be sure your setup meets GPIB protocol and
interface requirements. Then do the following procedures.
Connect the Oscilloscope to the GPIB. To connect the oscilloscope, plug an IEEE
Std 488.2-1987 GPIB cable (available from Tektronix, part number
012-0991-00) into the GPIB connector on the oscilloscope rear panel and into
the GPIB port on your controller or hardcopy device (see Figure 3–51).
Digitizing Oscilloscope (Rear Panel)
Controller
Figure 3–51: Connecting the Digitizing Oscilloscope to a Controller
Select the GPIB Port. To select the GPIB port, press SHIFT
➞ UTILITY ➞
System
(main) ➞ I/O (pop-up) ➞ Port (main) ➞ GPIB (pop-up).
Configure the GPIB Port. To set the GPIB mode and address, do the following:
Press SHIFT UTILITY
➞ System (main) ➞ I/O (pop-up) ➞ Port (main) ➞
GPIB (pop-up) ➞ Configure (main) ➞ Talk/Listen Address, Hardcopy (Talk
Only), or Off Bus
(side). (See Figure 3–52.)
Talk/Listen Address configures the port for normal controller based system
operation. Use the general purpose knob to define the address.
To Set Up for Remote
Operation