EasyManua.ls Logo

Tektronix TDS 420A - Page 101

Tektronix TDS 420A
433 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–87
CURSor:PAIred:POSITION<x>
Sets or queries the vertical bar (time) position of the first or second paired cursor.
x is either 1 or 2 and refers to the first or second cursor.
The CURSor:VBArs:UNITS command specifies the units for these cursors.
In extended-acquisition-length mode, the paired cursor position must be within
the waveform record (as opposed to the entire extended acquisition length) for
any change to happen.
Cursor
CURSor:FUNCtion, CURSor:VBArs:UNITS
CURSor:PAIred:POSITION <NR3>
CURSor:PAIred:POSITION?
CURSor
:
PAIred :
?
<Space> <NR3>
POSITION <x>
<NR3> specifies the cursor position in the units specified by the
CURSor:VBArs:UNITS command.
The position can appear in units of seconds, 1/seconds (hertz), or video line
numbers (with option 05). On the TDS 400A, position can also appear in units of
clocks or 1/clocks.
CURSOR:PAIRED:POSITION1 9.00E-6
specifies the first paired cursor is at 9 ms.
CURSOR:POSITION2?
might return 1.00E-6, indicating that the second paired cursor is at 1 ms.
CURSor:PAIred:POSITION<x>Pcnt
TDS 400A Only
Sets or queries the horizontal position of the paired cursors (x is either 1 or 2) in
units of % of record length.
Group
Related Commands
Syntax
Arguments
Examples

Table of Contents