Maintenance
6-- 32
TDS1000 and TDS2000 Series Digital Storage Oscilloscopes Service Manual
See
boot up
screen?
Errors?
Front panel
problems?
Run probe check
on each channel
Put scope or
frequency counter
on probe comp
Replace main
board
1 kHz?
Verify LCD
connector and
cable values
Good
voltages?
Replace main
board
Replace display
or display cable
Yes
No
Probe
check Ok?
Replace main
board
Connect
external trigger
to probe comp
Set trigger
source to EXT
Press set 50%
Trigger
indicator ON?
No
Yes
Replace main
board
Good
Oscilloscope
Yes
No
Yes
Yes
No
Yes
No
Yes
Turn on power
Does screen
lighten?
Yes
From Figure 6-21
To Figure 6-21
No
A
B
To Figure 6-20
C
To Figure 6-20
D
No
No
Figure 6--19: Oscilloscope troubleshooting tree (1 of 3)

Table of Contents

Other manuals for Tektronix TDS2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS2014 and is the answer not in the manual?

Tektronix TDS2014 Specifications

General IconGeneral
Channels4
Bandwidth100 MHz
Sample Rate1 GS/s
Record Length2.5k points
Vertical Resolution8 bits
Vertical Sensitivity2 mV/div to 5 V/div
Timebase Range5 ns/div to 50 s/div
Display Resolution320 x 240 pixels
Display5.7 inch color LCD
Trigger TypesEdge, Video
InterfaceUSB
Input Impedance1 MΩ
Power Requirements100 to 240 V