Maintenance
TDS1000 and TDS2000 Series Digital Storage Oscilloscopes Service Manual
6-- 45
Clearing the Error Log. To clear the error log after fixing a problem, follow these
steps:
1. Enable the Service menus as described on page 5--3.
2. Push the UTILITY button and push the Service option button to display the
Service menu.
3. Push the Service Diag. option button to display the Diagnostics menu.
4. Push the Clear Error Log option button to clear the error log of past
failures. This also sets the Power Cycle count to zero.