EasyManuals Logo
Home>Tektronix>Test Equipment>TDS3032B

Tektronix TDS3032B User Manual

Tektronix TDS3032B
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #120 background imageLoading...
Page #120 background image
Horizontal Controls
3-38
TDS3000B Serie s User Manual
Delay Interactions. The maximum delay setting is a function of the
time-base setting and acquisition resolution. If you set a large
positive or negative delay, that amount of delay might be automati-
cally reduced if you make the following additi onal control cha nges:
H Change to a faster time-base setting
H Change from Fast Trigger to Normal acquisition resolution
If the delay reduction occurs, it may cause the waveform horizontal
position to shift.
Negative Delay. You can select up to ten divisions of negative delay.
At the fastest time base settings, you can use negative delay to see
more of the wave form t hat occurs before the trigger point.
Roll Mode Display. To obtain a rolling display similar to a strip-chart
recorder, turn off zoom and delay, select the Auto trigger mode, and
set the horiz ontal SCALE cont rol t o 40 ms/div or slower. Subsequent
changes to the horiz onta l SCALE cause the roll-mode display to
erase and restart.

Table of Contents

Other manuals for Tektronix TDS3032B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS3032B and is the answer not in the manual?

Tektronix TDS3032B Specifications

General IconGeneral
BrandTektronix
ModelTDS3032B
CategoryTest Equipment
LanguageEnglish

Related product manuals