Specifications
TDS3000B Series Service Manual
1-5
Table 1- 1: Specifications (cont.)
Horizontal
Acquisition
(horizontal) resolution
Normal
(10,000 point record)
Fast trigger
(500 point record)
Maximum 700 w
ms
s 3,600 wfms
s
acq rate, typical
,
Sample rate range TDS301xB TDS302xB,
TDS303xB
TDS304xB--
TDS306xB
Normal
100 S/s to
1.25 GS/s
100 S/s to
2.5 GS/s
100 S/s to
5GS/s
Fast trigger
5 S/s to
1.25 GS/s
5 S/s to
2.5 GS/s
5 S/s to
5GS/s
Seconds/division
range
4ns/divto
10 s/div
2ns/divto
10 s/div
1ns/divto
10 s/div
n Sample rate and
delay time accuracy
±20 ppm over any ≥1 ms time interval
Trigger
External trigger input,
typical
TDS30x2B:
1MΩ in parallel with 17 pF, TekProbe compatible
TDS30x4B:
1MΩ in parallel with 52 pF, not TekProbe compatible
External trigger Overvoltage category Maximum voltage
maximum voltage
CAT I Environment
(refer to page 1--11)
150 V
RMS
(400 V
pk
)
CAT II Environment
(refer to page 1--11)
100 V
RMS
(400 V
pk
)
For steady-state sinusoidal waveforms, derate at 20 dB/decade above
200kHzto13V
pk
at 3 MHz and above
External trigger maxi-
mum floating voltage
0 V from chassis (BNC) ground to earth ground
30 V
RMS
(42 V
pk
) only under these conditions: no signal voltages
>30 V
RMS
(>42 V
pk
), all common leads connected to the same voltage,
no grounded peripherals attached
n Edge trigger Source Sensitivity
sensitivity
Any channel, DC coupled
≤0.6 div from DC to 50 MHz,
increasing to 1 div at oscilloscope
bandwidth
Edge trigger Source Sensitivity
sensitivity, typical
External trigger
200 mV from DC to 50 MHz,
increasing to 750 mV at 300 MHz
External/10 trigger 500 mV from DC to 50 MHz,
increasing to 3 V at 300 MHz