26
d. TECHNICAL SPECIFICATIONS
Manufacturer TENEX HEALTH
Model TX
Classication (IEC 60601) CLASS I
PARAMETER SPECIFICATION
Electrical
INPUT VOLTAGE 100 to 240 VAC @ 50/60 Hz
MAXIMUM CURRENT 1.2 Amps
LINE FUSES 5A T (Slow-blow)
MAX. OUTPUT VOLTAGE (CUTTING) 250 Volts
MAX. OUTPUT VOLTAGE (COAGULATION) 15 Volts at 75 ohm load (Fixed output)
MAX. OUTPUT VOLTAGE (FOOT PEDAL) 5 Volts
Irrigation
FLUID DELIVERY Pressure fed: regulated compressor
OPERATING PRESSURE 4 PSI nominal
VALVE TYPE Solenoid-actuated Pinch Valve
CONTROL Foot Pedal
Cutting
TX MicroTip TYPE Sterile, Disposable Ultrasonic
OPERATING FREQUENCY 25–28 kHz
CONTROL Foot Pedal
SYSTEM PRIMING Automated Prime Cycle
Available Power Delivery
LOW 100% delivered at 50 ms ON and 50 ms OFF
MEDIUM 100% delivered at 75 ms ON and 25 ms OFF
HIGH 100% delivered Continuously
DUTY CYCLE (USER-CONTROLLED) 15s/45s (15 seconds ON, 45 seconds OFF at maximum
power)
Aspiration
ASPIRATION PUMP Peristaltic, Low Pulsation
AVAILABLE MAX. VACUUM LEVEL User selectable: Low/Medium/High (100/300/500 mmHg)
AVAILABLE FLOW RATE User selectable: Low/Medium/High (10/20/30 cc/min)
CONTROL Foot Pedal
Coagulation
TYPE Biploar Generator
OPERATING FREQUENCY 315kHz
MAXIMUM POWER 3.4 Watts at 75 ohm load (Fixed output)
ACCESSORY TYPE Bipolar Forceps or Pencil (user-provided accessory)
CONTROL Foot Pedal