f. COAGULATION FUNCTION
The COAGULATION Panel provides access to the parameter settings for the Coagulation
function. Controlled by depressing the Foot Pedal, COAGULATION sends an isolated 315 kHz
power signal to a user-provided bipolar cautery accessory to cauterize small blood vessels.
When COAGULATION is selected, all other functions are automatically set to OFF. Depressing
the Foot Pedal will activate the cautery accessory and an audible TONE will sound. When the
Food Pedal is released, the cautery accessory and audible tone will turn OFF.