6. Settings
15
6.1 Selecting fuel
1 Select Fuel in main menu ➀ and confirm selection
using .
- The set fuel flashes in the display ➁.
2 Select required fuel via or and confirm selection using
.
- Selected fuel is accepted and the analyser changes to the
measurement menu.
6.2 Setting display sequence
The sequence in which the parameters are shown in the display
can be selected as required. In this way, frequently used
parameters can be combined and then called up quickly
1 Select View in the main menu and confirm selection
with .
- Line1 flashes in the display and the set parameters and units
for line 1 and 2 are shown.
2 Use to select required parameter for line 1.
3 Use to change to Line 2 and select required parameter
for line 2 with .
4 Set up additional parameters for lines 3 to 12. Repeat
procedure as described in step 3.
5 Confirm settings with .
- Selected settings are accepted and the analyser changes to
the measurement menu.
Selectable parameters:
O
2
Oxygen level
Flue gas temperature (FT)
Ambient air
temperature (AT)
Differential temperature
(FT - AT)
EFF
NET
Efficiency referred to H
u
EFF
GROS
Efficiency referred to H
o
rat Ratio
E
AA
Excess Air (EXA)
CO
2
Carbon dioxide
CO Carbon monoxide
uCO Carbon monoxide
undiluted
Air ratio
P
1
Pressure (0 to 40hPa)
P
2
Pressure (0 to 200hPa)
td Dew point
1
CO
*
CO maximum
CO CO ambient
1
(testo 325XL only)