6. Measurement examples
39
Messprogramm konfiguriert
-
Overview of programmed long-term measurement appears.
8
Confirm measurement program with .
-
Measurement program takes effect and is configured:
9
Press function key .
-
Long-term measurement begins:
The symbol shows which program is currently running.
Long-term measurement starts with fresh air/rinse phase:
Fresh air time between measuring cycles:
Minimum and maximum measuring and fresh air cycles:
Measuring cycles: 2 min to 240 min (4 h)
Fresh air cycles: 5 min to 1440 min (24 h)
10
ends the programmed measurement ahead of time.
- The measuring program remains activated (indicated by the symbol ).
- Generally, there is a rinsing phase of 2 minutes after the measuring program is
finished (pump slow).
6.2.3 Restarting measuring program
Press function key.
- Measuring program starts again.
6.2.4 Deleting programming in instrument
1
Select –> –> –> .
Measuring program is only deactivated and is not deleted.
2
Select –> –> .
- The last measuring program activated is shown.
3
Press the function button to reactivate.
InfoProgram
Memory
DeleteProgram
Memory
Start
Stop
Start
Fresh aair rrinse ZZeroing 11 ZZeroing 22 EExposure tto ggas
t in min
Constant 6min
Measure
0 1 2 3 4 5 6
Rinse FFresh aair ZZeroing 11 ZZeroing 22 EExposure tto ggas
t
Set fresh air time
Measure
Rinsing is ended once
rinse limits are reached.
Measuring program in progress