EasyManua.ls Logo

TESTO 350 S - Restarting a Measuring Program; Deleting Programmed Settings

TESTO 350 S
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6. Measurement examples
39
Messprogramm konfiguriert
-
Overview of programmed long-term measurement appears.
8
Confirm measurement program with .
-
Measurement program takes effect and is configured:
9
Press function key .
-
Long-term measurement begins:
The symbol shows which program is currently running.
Long-term measurement starts with fresh air/rinse phase:
Fresh air time between measuring cycles:
Minimum and maximum measuring and fresh air cycles:
Measuring cycles: 2 min to 240 min (4 h)
Fresh air cycles: 5 min to 1440 min (24 h)
10
ends the programmed measurement ahead of time.
- The measuring program remains activated (indicated by the symbol ).
- Generally, there is a rinsing phase of 2 minutes after the measuring program is
finished (pump slow).
6.2.3 Restarting measuring program
Press function key.
- Measuring program starts again.
6.2.4 Deleting programming in instrument
1
Select –> –> –> .
Measuring program is only deactivated and is not deleted.
2
Select –> –> .
- The last measuring program activated is shown.
3
Press the function button to reactivate.
InfoProgram
Memory
DeleteProgram
Memory
Start
Stop
Start
Fresh aair rrinse ZZeroing 11 ZZeroing 22 EExposure tto ggas
t in min
Constant 6min
Measure
0 1 2 3 4 5 6
Rinse FFresh aair ZZeroing 11 ZZeroing 22 EExposure tto ggas
t
Set fresh air time
Measure
Rinsing is ended once
rinse limits are reached.