7 Maintaining the product
51
7 Maintaining the product
7.1. Servicing interval
Testo recommends having the measurement system checked once
a year by Testo Service.
7.2. Cleaning the testo 380 and testo 330
> In order to prevent damage to the measurement system, the
measurement system must be disconnected from the mains
when being serviced and cleaned.
> If the housing of the testo 380 and the testo 330 is dirty, clean
it with a damp cloth. Do not use any aggressive cleaning agents
or solvents! Mild household cleaning agents and soap suds
may be used.
> Clean ventilation slots, gas outlets, fresh air inlet and dilution air
inlet with a vacuum cleaner. Do not blow out with compressed
air.
7.3. Draining the condensate container
The condensate consists of a weak mix of acids. Avoid skin
contact. Make sure that the condensate does not run over
the housing.