What to do if Thermo Scientific Laboratory Equipment can't reset pressure sensor because pressure is out of range?
- RrherreraAug 27, 2025
If the pump pressure is not down to zero during the self test, wait until the pressure is down and repeat the test. Verify that the cable for the system pressure transducer is properly connected to the P-Sys connector for the loading pump on the interior front panel. Open and close the purge valve if necessary to reduce the pressure.