Why Thermo Scientific Laboratory Equipment cannot regulate flow?
- LLisa CarlsonAug 25, 2025
If the NC pump cannot build up sufficient pressure to deliver the selected flow rate, consider these potential causes and solutions: * Ensure the solvent reservoir is full. * Purge the pump to remove any trapped air. * Check for leaks in the system; find and eliminate any sources of leakage, tightening connections as needed. * Replace the frit in the solvent supply line filter if it's clogged. * Replace the inlet valve cartridge if it is blocked.