Thermo Scientific HyPerforma G3Pro Universal Bioreactor Controller User’s Guide | 14
Chapter 1 | G3Pro overview
Optional Universal Power Suppy (UPS)—A backup power unit
(APC Smart UPS). It supplies power for a given amount of time
(approximately 20 minutes) in the event of power failure which could
be catastrophic during a process event.
HMI/MonitorSupplies power for the computer monitor.
Vessel Adapter BoxThis connector provides control for large
S.U.B. electric heaters. Used in conjunction with the Vessel Heater
In connector when operating without a TCU.
Vent Filter HeatersThese connectors provide a paired
connection for each of the two (2) vent filter heaters. Power is
supplied through the receptacles. Note: Because the voltage used
for the unit is carried through to the vent filter heater receptacles,
you must ensure that the vent filter heaters are rated for the
appropriate voltage.
Main Power Switch (On)—Turns main power on and o.
Ground Stud–Used to connect a permanent true earth bonding
conductor (see section 2.4.2).
Power-indicator/Status lightIlluminates green when power is
on.