EasyManuals Logo

Thermo Orion Aplus Series User Manual

Thermo Orion Aplus Series
131 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #20 background imageLoading...
Page #20 background image
Orion Benchtop pH/ISE Meter Instruction Manual
Calibration and Measurement of pH
16
Temperature Compensation
pH measurements on the Orion 410Aplus are made with Automatic
Temperature Compensation. pH measurements on the Orion 420Aplus,
520Aplus, 525Aplus, 710Aplus, 720Aplus, and 920Aplus may be made
with either Automatic or Manual Temperature Compensation.
For Automatic Temperature Compensation, an ATC probe must be used.
Plug in the ATC probe and the meter will display temperature corrected pH
results in the main display.
For Manual Temperature Compensation with Orion 520Aplus, 525Aplus,
720Aplus, and 920Aplus disconnect the temperature probe. Temperature
values can be entered manually by pressing set temp. Temperature
corrected pH values, based on the manually entered temperature, will be
displayed in the main field.
For Manual Temperature Compensation with Orion 420Aplus and
710Aplus disconnect the temperature probe. Temperature values can be
entered manually by pressing or while in measure mode. The value
will be displayed in the lower field. Temperature corrected pH values based
on the manually entered temperature will be displayed in the main field.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Orion Aplus Series and is the answer not in the manual?

Thermo Orion Aplus Series Specifications

General IconGeneral
BrandThermo
ModelOrion Aplus Series
CategoryMeasuring Instruments
LanguageEnglish