Orion Benchtop pH/ISE Meter Instruction Manual
Special Features
38
Cal Edit
Prompt Key Press/Action
1. Press 2nd
then cal edit.
2. STD 1 OK? The meter displays the current value
of standard one. Press yes if okay or
no to edit.
3. If no selected: Press 1 or 2 then press yes.
1-REMEAS 2-EDIT
4. If 1 selected: Place electrode(s) in the standard
READY ENTER and wait for a stable display.
VALUE Enter the value of the standard and
press yes.
5. If 2 selected: Press 1 or 2 then yes.
1-CHANGE
2-DELETE
6. If 1 selected: The meter prompts for the new
ENTER VALUE value. Enter value and press yes.
7. If 2 selected: The meter deletes the calibration
STD 2 OK? point and continues to the next
calibration point. After presenting all
calibration points for editing, the
meter returns to measure mode.