Table II Operator Assistance Codes
Error
Code Problem Probable Remedy
E-20 out of range 1. If this occurs when electrode(s)
pH,mV, are out of solution, code will
Rel mV, or conc. disappear when electrode(s) are
returned to solution.
2. Verify electrode(s) are
properly connected.
3. Sample may be out of range.
Check buffer or standard.
4. Check slope value and the
isopotential point setting.
5. Recalibrate using fresh buffer
or standards.
6. Check that the ATC probe is
connected properly and make sure
temperature is within -5.0
to +105˚C.
E-21 ABR error Electrode voltage being measured
pH autocalibration is > ±0.5 pH units from the
error nominal value for the pH buffer,
therefore the meter does not
recognize it as an autobuffer.
1. Verify buffer as 1.68, 4.01,
7.00, 10.01, or 12.46. For Orion
410Aplus and 420Aplus verify
buffer as 4.01, 7.00, or 10.01.
2. Press any key and recalibrate using
fresh buffer. .
3. Clean electrode(s) and refill
reference electrode. Recalibrate.
4. Perform a manual calibration.
Some electrode(s) may operate
outside of the range for pH
autocalibration. This may be a sign
for an electrode aging and
is normal.
Orion Benchtop pH/ISE Meter Instruction Manual
Troubleshooting
70

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Orion Aplus Series and is the answer not in the manual?

Thermo Orion Aplus Series Specifications

General IconGeneral
BrandThermo
ModelOrion Aplus Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals