2. Outline and Specifications
ARV-310 2-9
2-4-2 Other Operating Parts
No.
Name Function
POWER switch Used for turning ON/OFF the power.
EMERGENCY
button
Used for stopping operation in emergency.
Press to stop the container rotation and the
vacuum pump operation. Air in the vacuum
chamber is released to ambient pressure.
In addition, all lamps (④、⑤、⑬) and “ALM”
indications of all indicators (①、③、) on the
control panel start blinking with a buzzer
sound.
This button is locked when pushed in
(push-lock type).
For restarting, turn the button C.W. to release
the lock, turn OFF (O) the POWER switch,
and turn it ON (I).