5. Maintenance
ARV-310 5-13
5-5 Error Display
The Thinky ARV-310 incorporates a lid sensor, vibration
sensor and revolution sensor. If any error occurs during
revolution, a relative message is displayed on the control
panel and operation stops accordingly.
Major possible causes and measures for correction are listed
below.
If the cause of an error is clear, remove it.
If the causes or the contents of the error are not clear, contact
the dealer where you purchased the unit or Thinky direct.
For details of the contact, refer to “Introduction” in this Manual.
Error Display
Cause Measures
Operation was started with
the lid open.
Err1
The lid was opened during
operation.
Close the lid to operate.
Abnormal vibration occurred
abruptly during operation.
Check if the container
has not been detached,
flown away or any parts
have been damaged.
Err2
The vibration value
exceeded the rated value.
Adjust the counter
balance.
Err3 - Err6
Err10
Revolution speed does not
increase after the operation
has started.
Contact the dealer
where you purchased the
unit or Thinky direct.