5. Maintenance
ARV-310 5-5
When the oil is contaminated, change the oil.
For changing oil, refer to “5-3-2 Oil Change”.
If the oil level has lowered to the center of the oil check window,
add oil to the level of two-thirds (2/3) of the window.
For replenishing oil, refer to “5-3-2 Oil Change”.
2. Time to Reach Vacuum Decompression
Perform vacuum decompression to check the period of time
required to finish decompression referring to the followings as
a guideline.
Time required for decompression (guideline):
Within 1 min. (60 sec.) to 2.66 kPa (about 20 Torr), and
within 3 min. (180 sec.) to 530 – 660 Pa (about 4 – 5 Torr)
If decompression operation does not finish within the above
guideline, change the oil.
For changing oil, refer to “5-3-2 Oil Change”.
After the oil change, perform decompression for about 15
minutes to check if decompression operation finishes within
the above guideline.
Depending on materials that got mixed in the oil, changing oil only
once may not be sufficient. In such case, repeat changing oil two or
three times.
Odor of solvents, etc. is difficult to get rid of. It cannot be completely
eliminated.