7
DIGITAL TORQUE WRENCH TESTER DOTE4-G
5-7. Statistical Processing Function, Number of samples, Max-Min-Mean Value
#1 Use ▼/▲ keys and show the last result of the section for statistical processing.
#2 Press “STAT” key to show “Stt” on display. And use ▼/▲ keys to show the rst result of the section.
Press “STAT” key and the sampling number “n”, maximum value “HI”, minimum value “Lo” and mean
value “Av” in the specied data range are displayed in order.
5-8. Auto Zero Adjustment
With no load on the tester, Auto Zero function is performed automatically by press C” key or turn the
power OFF/ON switch.
“Err9” on display
When loading torque exceeds more than 5% of the tester's maximum torque range, “Err9” is displayed.
When “Err9” is displayed, press “C” key with no load on the tester. “Err9” message disappears.
If “Err9” message is still on the display, turn off and on again.
Torque sensor or the electric circuit board may have any problem if “Err9” shows again.
5-9. Over-Torque Alarm
When loading torque exceeds 110% of its maximum torque range, the display ashes as a warning.
5-10. Error Display
DOTE4-G has a self-diagnosis function. When a problem occurs, any of the error messages from Err1
to 9 is displayed.
For any other error messages, refer to "Section #10. Error Message"