23
DIGITAL TORQUE WRENCH TESTER DOTE4-G
Error Message Indication Solution
Err1 – 5
Operation key is continuously
pushed.
Turn off DOTE4-G once and turn it on without touching any keys.
If Err disappears, it should work properly.
If Err does not disappear, it needs to be repaired.
Please contact TOHNICHI or your nearest distributor.
Err8 CPU / Memory error.
It needs to be repaired. Please contact TOHNICHI or
your nearest distributor.
Err9
Malfunction of the torque
sensor or circuit board.
Push C key at no loading condition.
If Err 9 disappears, it should work properly.
If Err 9 does not disappear, it needs to be repaired.
Please contact TOHNICHI or your nearest distributor.
10. Error Message