11
DIGITAL TORQUE WRENCH TESTER DOTE4-G
#3 Data length setting
Use ▲/▼ keys and select “7bit” or “8bit”. Press “STAT” key to enter and got to next setting,
“Parity Setting”. Press “MD” key to skip the setting. Press “C” key to go back to the measurement mode.
Factory Default Setting : 8bits
#4 Parity setting
Use ▲/▼ keys and select “nonE” for none, “EVEn” for even number or “odd” for odd number.
Press “STAT” key to enter and go back to the measurement mode. Press “MD” key to skip the setting.
Press “C” key to go back to the measurement mode.
Factory Default Setting : None
Memory mode setting
Set a measured data memory capacity.
“1000”: 1,000 data memory
“99”: 99 data memory, communication format compatible with previous DOTE3-G model
Factory Default Setting : 1000
Use ▲/▼ keys and select 1000 or 99. Press “STAT” key to go conrmation screen.
Press “MD” key to skip the setting. Press “C” key to go back to the measurement mode.
Press “STAT” key to conrm. Memory mode is changed and all stored data is cleared.
When memory mode is changed, existed memory is cleared. Press “C” key to go back.
Default setting
Use ▲/▼ keys to select “dFLt-Y”. Press “STAT” key to reset to the factory default settings.
Press “MD” key to skip the setting. Press “C” key to go back to the measurement mode.