3. Various Settings
59
3.12 Average Measure
The average measurement can be set.
Display Method
3.1.2 Data/Setting Item Display
When measuring varying-luminance light source, use this function. Measurement result in
Average measure is the average value of 5 times measurement and, measurement interval
is about 1 second.
1
Press the [CHANGE] button.
2
Press the [ROTATION] button and select the setting contents.
Each time the button is pressed, the display will change as shown below.
Single Average Single ……
3
After selection is finished, press the [ENTER] button.
3.13 Maintenance Recommended Display
The maintenance recommended display is displayed after buying the or after one year has elapsed
since calibration conducted by TOPCON TECHNOHOUSE factory. BM-7AC
Display Method
3.1.2 Data/Setting Item Display
ON : A buzzer will sound and the display will not advance to the next screen until any
button is pressed.
OFF : The message is displayed but the buzzer does not sound and it is possible to
advance to the next screen.
The procedure for making changes is as described below.
1
Press the [CHANGE] button.
2
Press the [ROTATION] button and select the setting contents.
Each time the button is pressed, the setting will change between ON/OFF.
3
After selection is finished, press the [ENTER] button.
Maintenance Display
*ON
Average Measure
Single