4. Communication With a PC
65
4.2.9 R Command
The Correction factor in the instrument is read out.
R [n] : n is the number of the correction factor to be read. 1 to 15
PC Command BM-7AC
R15 + (CrLf)
OK + (CrLf)
Correction Factor
END + (CrLf)
4.2.10 W Command
The correction factor is stored in the instrument.
W [n] : n is the number of the correction factor to be stored. 1 to 15
PC Command BM-7AC
W10 + (CrLf)
OK + (CrLf)
4.2.11 CF Command
The correction factor in the instrument is cleared.
CF [n] : n is the number of the correction factor to be cleared. 1 to 15
PC Command BM-7AC
CF10 + (CrLf)
OK + (CrLf)