2
Pinhole
Cross hairs glass
Standard wearing distance: 12mm, 13.75mm, 16mm, 18mm, 20mm
Convergence: Near-point distance 40/67cm (Minimum pupillary distance to
enable convergence at the near point 40cm: 53mm)
Forehead rest adjustment range: 15mm
1-3-2 Other specifications
Horizontal adjustment: 4° or more in right and left
Tilt for near point: 30°
Near-point illumination: White LED
Input of objective data/CL data: Input the measurement data of RM or CL into CV.
Remote control of ACP-8/MC-3/PC-2: Remote control of ACP-8, MC-3 or PC-2 is possible.
Electric rating:
For Japan Power supply voltage: AC100V, Frequency: 50-60Hz
For export Power supply voltage: AC100-240V, Frequency: 50-60Hz
Power supply input: 90VA
Dimensions and weight:
Dimensions: 294 ~ 328mm (W) × 115mm (D) × 261mm (H)
Weight: 4kg
Types of protection against electric shocks: Class I equipment
Degree of protection against electric shocks: Type B applied part
1-4 Target Users of This Repair Manual
Only the service engineer having experience in assembling and adjusting ophthalmic instruments must repair
this instrument. For others, it is recommended to take the repair training lessons of TOPCON
CORPORATION.