OPHTHALMIC UNIT
IS-100
Rev. 6 - November 2019Rev. 6 - November 2019
6-4
USER MANUAL
The iron should not be on the surface of the tabletop for more than 5
seconds. In addition, during ironing, the iron must be kept in constant
motion on the paper towel.
4. Remove the paper towel and dry the surface.
5. Repeat the operation as many times as necessary.
It is not possible to repair with this procedure scratches that have
penetrated the protective surface of the tabletop.
6.3.2. Lifting motor of the chair not operational
If the manoeuvre of the lifting motor is not operational, it may be due to the
activation of the thermal protection due to an overload, or a continuous job
without respecting the work cycle.
The work cycle allows 2 minutes of work within a range of 18 minutes.
In order to solve this problem, carry out the following checks:
y Wait for the reset: if the fault is due to an overload of the lifting group, wait
until the temperature drops.
y Check the corresponding fuses.
BEFORE REPLACING THE FUSES, DISCONNECT THE UNIT FROM
THE POWER SUPPLY.
y Checkthatthesupplyvoltagematchesthespecicationsoftheunit.
y Check the correct connection of the power supply wiring.
If, after these checks, the fault persists, contact the technical support service.