12. ERROR DISPLAY
67
If errors still persist after attempting to clear them, contact Topcon or your dealer.
Error Display Description/Countermeasure
E-05 Turn the power for the instrument off, and then turn it back on.
E-51,55
Internal error for the RC-50
Transmission not possible with the RL-SV1S.
Remove and replace the dry cell batteries from the RC-50.
E-56
Wireless function error for the RL-SV1S.
Unable to transmit with the RC-50.
Turn the power for the instrument off, and then turn it back on.
E-65
Internal transmission error for the RL-SV1S.
Turn the power for the instrument off, and then turn it back on.
E-70’s
Slope function error.
Turn the power for the instrument off, and then turn it back on.
E-80’s
Leveling incomplete.
Turn the power for the instrument off, and then turn it back on in stable
locations.
E-99
Internal memory error for the RL-SV1S.
Turn the power for the instrument off, and then turn it back on.